Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions...
Medtronic Powered Surgical Solutions began a recall of Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial... on . The reason FDA records is "Cutting flute geometry of surgical dissecting tools may be outside of established tolerance. This may lead to reduced cutting performance, inability to perform desired dissection, increased risk of overheating, and...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2156-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Powered Surgical Solutions
- Product
- Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General Surgical and Plastic Surgery applications for transection, osteotomy, graft harvesting, bone shaping, laminectomy, entry hole, suture hole, and midface advancement.
- Product code
- HBB Motor, Drill, Pneumatic
- Reason for recall
- Cutting flute geometry of surgical dissecting tools may be outside of established tolerance. This may lead to reduced cutting performance, inability to perform desired dissection, increased risk of overheating, and potential need for revision surgery.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Firm notified consignees by Urgent: Medical Device Recall letter sent May 30, 2008. The letter requires consignees to immediately compare their inventory of products against the list provided and quarantine any unused dissecting tools. The letter also states that Medtronic would provide replacement product to match the quantity and type returned by the consignees' institution.
- Quantity in commerce
- 1,306 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Lot numbers: 0000710253, 0000710254, 0000710256, 0000710257, 0000710258, 0000710266, 0000710267, 0000710271, 0000710272, 0000710274, 0000710275, 0000710276, 0000710277, 0000710278, 0000710279, 0000710280, 0000710281, and 0000710288