Z-2159-2008

10% Buffered Formalin Phosphate; Certified For laboratory and manufacturing use only, not for drug, food or...

Thermofisher Scientific Oy / initiated 2008-05-01 / Terminated / root cause: manufacturing process

ThermoFisher Scientific Co., LLC began a recall of 10% Buffered Formalin Phosphate; Certified For laboratory and manufacturing use only, not for drug, food or household use. Do not Transfer to an unmarked... on . The reason FDA records is "Assay tests on retention sample of 10% Buffered Formalin Phosphate showed it to be 3.417%, which is below the specification of 4-5%". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2159-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Thermofisher Scientific Oy
Product
10% Buffered Formalin Phosphate; Certified For laboratory and manufacturing use only, not for drug, food or household use. Do not Transfer to an unmarked container. Catalog numbers: SF100-4, SF100-20 Fisher Scientific, Fail Lawn, NJ 07410
Product code
IFP Formalin, Neutral Buffered
Reason for recall
Assay tests on retention sample of 10% Buffered Formalin Phosphate showed it to be 3.417%, which is below the specification of 4-5%.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Urgent: Product Recall notification letters were sent by US mail on May 1, 2008. The letters asked that those in receipt of the product check their stock for this product, destroy any of the material that remains from the referenced lot, and then contact their local customer service center for a replacement. The letter also states that the appropriate corrective measures have already been instituted to prevent a recurrence.
Quantity in commerce
1052 x 4 liters, 564 x 20 liters
Distribution
Worldwide Distribution --- including USA and Canada.
Product codes and lots
Lot 075923