Z-2168-2008

Suction Jar Lid for Medela Vario 18 Vacuum Pump used with the Bravo pH Monitoring System Medtronic Part...

Medtronic Neuromodulation / initiated 2008-04-25 / Terminated / root cause: manufacturing process

Medtronic Neuromodulation began a recall of Suction Jar Lid for Medela Vario 18 Vacuum Pump used with the Bravo pH Monitoring System Medtronic Part Number: 9043K2101. (9043K2111 is the OUS version) The... on . The reason FDA records is "The Medela Vario 18 Vacuum Pump used with the Medtronic Bravo pH Monitoring System. The Vacuum Pump is used to provide the vacuum pressure during Capsule attachment. Some Medela Vario 18 Vacuum Pumps were provided with...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2168-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Suction Jar Lid for Medela Vario 18 Vacuum Pump used with the Bravo pH Monitoring System Medtronic Part Number: 9043K2101. (9043K2111 is the OUS version) The Vacuum Pump is used to provide the vacuum pressure during Capsule attachment. Correct Overflow lid: Medela CH-6341 Baar Switzerland Art. No. 077.0440, Small lid with conical patient connection 6-10mm and overflow protection device.(Incorrect overflow lid is Medela P/N: 077.0450)
Product code
BTA Pump, Portable, Aspiration (Manual Or Powered)
Reason for recall
The Medela Vario 18 Vacuum Pump used with the Medtronic Bravo pH Monitoring System. The Vacuum Pump is used to provide the vacuum pressure during Capsule attachment. Some Medela Vario 18 Vacuum Pumps were provided with an incorrect suction Jar Lid that did not have the correct conical connection for the Vacuum Tube.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Affected customers contacted by letter "Medtronic Urgent: Medical device Recall April 2008" that indicates that some Medela Vario 18 Vacuum Pumps were provided with an incorrect Suction Jar Lid that did not have the correct Conical Connection for the Vacuum Tube and the method to obtain a replacement part if needed. The letter requires customers to locate the Vacuum Pump serial number and compare it against the Serial Number on the Reply Form provided. Next customers are to review the Suction Jar Lid and check the part number. "Reply form and Instruction for Receiving Replacement Lid " included. Contact Medtronic 1-800-707-0933 for assistance.
Quantity in commerce
1500 pumps (20 affected)
Distribution
USA including states of OH, DE, CT, KY, AR, NY, MA, CA, GA, NC, MI, WI, TX, PA, MD, and country of CANADA.
Product codes and lots
Serial #:  1163101 - 1163120