Ev3 Primus GPS Biliary Stent System 8 mm, 37 mm, B7, .40 mm, 5Fr, REF BXB35-08-37-75 Use before 2009-02...
Ev3, Inc began a recall of Ev3 Primus GPS Biliary Stent System 8 mm, 37 mm, B7, .40 mm, 5Fr, REF BXB35-08-37-75 Use before 2009-02 Sterile EO Ev3 4600 Nathan Lane North, Plymouth, MN... on . The reason FDA records is "A Primus GPS Biliary Stent System was mislabeled. The mislabeling resulted in a 8 mm diameter, 27 mm length stent being labeled as a 8 mm x 37 mm stent". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2170-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Ev3 Primus GPS Biliary Stent System 8 mm, 37 mm, B7, .40 mm, 5Fr, REF BXB35-08-37-75 Use before 2009-02 Sterile EO Ev3 4600 Nathan Lane North, Plymouth, MN 55442-2920 Intended as a palliative treatment of malignant neoplasms in the biliary tree.
- Product code
- FGE Stents, Drains And Dilators For The Biliary Ducts
- Reason for recall
- A Primus GPS Biliary Stent System was mislabeled. The mislabeling resulted in a 8 mm diameter, 27 mm length stent being labeled as a 8 mm x 37 mm stent.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The single affected consignee was visited by an ev3 representative to verify the use of the device or exchange it for an unaffected device. Information recorded on an ev3 Device Reconciliation Form.
- Quantity in commerce
- 1 Unit
- Distribution
- Nationwide Distribution --- state of NJ
- Product codes and lots
- Lot 2092998