Stem Trial, Triathlon Revision Instruments; Non-Sterile; 150 mm length Catalog Numbers: 5566-T-009...
Stryker Howmedica Osteonics Corp. began a recall of Stem Trial, Triathlon Revision Instruments; Non-Sterile; 150 mm length Catalog Numbers: 5566-T-009, 5566-T-010, 5566-T-011, 5566-T-012, 5566-T-013, 5566-T-014... on . The reason FDA records is "Triathlon TS Stems Trails may crack, potentially causing fragments of the plastic handle to fragment and break off completely". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2158-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stem Trial, Triathlon Revision Instruments; Non-Sterile; 150 mm length Catalog Numbers: 5566-T-009, 5566-T-010, 5566-T-011, 5566-T-012, 5566-T-013, 5566-T-014 5566-T-015, 5566-T-016, 5566-T-017, 5566-T-018, 5566-T-019, 5566-T-020 5566-T-021, 5566-T-022, 5566-T-023, 5566-T-024, 5566-T-025 Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 Used in Revision TKA.
- Product code
- HWT Template
- Reason for recall
- Triathlon TS Stems Trails may crack, potentially causing fragments of the plastic handle to fragment and break off completely.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Urgent Product Recall notification letters were sent to all Stryker Branch/Agency Manager/Quality contacts on May 14, 2008 via Federal Express. The letter requests that consignees examine their inventory and hospital consignment locations to identify the product; and retrieve all affected product lots and return it to their branch or agency warehouse for reconciliation. The letter also states that a Stryker Orthopaedics Customer Service rep should be contacted for alternative products available.
- Quantity in commerce
- Catalog 5566 T-009 to 025: 2625 US, 239 international
- Distribution
- Worldwide Distribution --- including USA and countries of Australia, Canada, Sweden, Germany, Spain, Portugal, United Kingdom.
- Product codes and lots
- All Lot Codes