Z-1902-2008
Levitronix CentriMag Back-Up Console Article Number: 201-30200; extracorporeal blood circulation system
Levitronix, LLC / initiated 2008-03-17 / Terminated / root cause: design
Levitronix, Llc began a recall of Levitronix CentriMag Back-Up Console Article Number: 201-30200; extracorporeal blood circulation system on . The reason FDA records is "Stops pumping: Interruption of CentriMag System support may occur when using a Valleylab Force FX-C Electrocautery Unit". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1902-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Levitronix, LLC
- Product
- Levitronix CentriMag Back-Up Console Article Number: 201-30200; extracorporeal blood circulation system.
- Product code
- KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
- Reason for recall
- Stops pumping: Interruption of CentriMag System support may occur when using a Valleylab Force FX-C Electrocautery Unit
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Levitronix issued an Urgent - Voluntary Device Correction letter on 3/17/08 mailed to the US distributor and requested to contact their accounts. The letter revises the warning in the CentriMag Console Manual. Levitronix is issuing revised manuals. On March 13, 2008 the firm emailed and mailed (FedEx) to the one Canadian Hospital and Health Canada was informed of the Voluntary Advisory Notice via email on Wednesday, March 19, 2008. Rest of the World: Wednesday, March 12, 2008 through Monday, March 17, 2008. The Notice was issued as a Voluntary Advisory Notice after review and follow up with the UK MHRA. On 7/24/08 the firm issued a revised Recall Letter to advise users not to use Valley Lab Force FX-C or SSE2L Electrosurgery device with the firm's CentriMag Blood Pumping System and include the significant risk to health. The firm revised their letter and Black Box Warning and reissued the letters to the hospital, surgeons, etc. on 7/24/08. On 2/5/09 the firm issued a "Follow-Up to Urgent Device Correction of 3/17/08 " to advise users of the software upgrade to be installed in the CentrigMag Primary software Version 3.0 and CentriMag Back Up console , software version 2.0. Site visits will be made by Levitronix between Feb 2009 and April 2009 , software upgrade installed and new Operating Manuals provided.
- Quantity in commerce
- 237 units
- Distribution
- Worldwide Distribution: USA, Canada, Switzerland, UK, Italy, France, Greece, Spain, Slovenia, Czech, Turkey, South Africa, Argentina, Chile, Panama, Singapore, Taiwan, Israel, and Saudi Arabia.
- Product codes and lots
- All serial numbers