Allez Spine Laguna Pedicle Screw System, Non-Conical Screw Size (diameter x length) 5 x 40mm, Winged, Catalog...
Allez Spine, LLC began a recall of Allez Spine Laguna Pedicle Screw System, Non-Conical Screw Size (diameter x length) 5 x 40mm, Winged, Catalog # LG-PW0540 on . The reason FDA records is "This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2071-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Allez Spine, LLC
- Product
- Allez Spine Laguna Pedicle Screw System, Non-Conical Screw Size (diameter x length) 5 x 40mm, Winged, Catalog # LG-PW0540
- Product code
- MNI Orthosis, Spinal Pedicle Fixation
- Reason for recall
- This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Allez Spine informed all parties who were in possession of the affected pedicle screws at the time of the initiation of this field action by phone on 11/08/2007 and requested the return of their inventory. All customers and consignees were contacted by the firm.
- Quantity in commerce
- 69
- Distribution
- Nationwide to AZ, CA, CO & UT