Z-2177-2008

Flexipet Manipulation pipette 300 micron. The device is used for the intracytoplasmic single sperm injection...

Cook Vandergrift, Inc. / initiated 2008-03-02 / Terminated / root cause: manufacturing process

Cook Vascular Inc. began a recall of Flexipet Manipulation pipette 300 micron. The device is used for the intracytoplasmic single sperm injection of oocytes on . The reason FDA records is "mislabeled - 300 micron flexipets labeled as 130 micron and vice versa". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2177-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Vandergrift, Inc.
Product
Flexipet Manipulation pipette 300 micron. The device is used for the intracytoplasmic single sperm injection of oocytes.
Product code
MQH Microtools, Assisted Reproduction (Pipettes)
Reason for recall
mislabeled - 300 micron flexipets labeled as 130 micron and vice versa
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The recalling firm issued an Urgent product Recall letter on 3/3/08. The letter informed the customers of the problem and that the pouch is correctly labeled. Customers were instructed to return product and notify any sub accounts. Please contact Cook Vascular at 1-724-845-8621 for assistance.
Quantity in commerce
120 boxes
Distribution
The product was shipped to physicians in CA, FL, GA, IN, NJ, NM, NV, WA , and WI.
Product codes and lots
Lot number N74173