Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2179-2008

Zimmer 4.5 Compression plate-broad, 10 hole, 167 MM length, stainless steel, non sterile, Manufactured at...

Zimmer GmbH / initiated 2008-06-09 / Terminated / root cause: manufacturing process

Zimmer Inc. began a recall of Zimmer 4.5 Compression plate-broad, 10 hole, 167 MM length, stainless steel, non sterile, Manufactured at Zimmer; REF . The product is used for... on . The reason FDA records is "An impurity in the metal may affect the strength of the plate or biocompatibility of the material with the body". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2179-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer 4.5 Compression plate-broad, 10 hole, 167 MM length, stainless steel, non sterile, Manufactured at Zimmer; REF 4945-10-01. The product is used for stabilization of fractures during the normal healing process.
Product code
HRS Plate, Fixation, Bone
Reason for recall
An impurity in the metal may affect the strength of the plate or biocompatibility of the material with the body.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Zimmer Thailand was notified via an emailed letter dated 6/9/08 to check for inventory at all customer accounts and to return the product. Contact Zimmer at 1-574-371-8028 for assistance.
Quantity in commerce
7
Distribution
International Distribution --- including Thailand.
Product codes and lots
Lot 60985554.