GE Centricity AW Suite software version 2.0 and 2.0.1 with Card IQ Xpress Pro or Plus Module; for diagnostic...
GE Healthcare Integrated IT Solutions began a recall of GE Centricity AW Suite software version 2.0 and 2.0.1 with Card IQ Xpress Pro or Plus Module; for diagnostic image analysis; GE Healthcare Integrated IT... on . The reason FDA records is "When using any vessel analysis protocol of the AW Suite Card IQ application, the software can reload saved tracking objects (Save State) incorrectly and display an incorrect vessel label over the restored images. This...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2183-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Centricity AW Suite software version 2.0 and 2.0.1 with Card IQ Xpress Pro or Plus Module; for diagnostic image analysis; GE Healthcare Integrated IT Solutions, Barrington, IL 60010
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- When using any vessel analysis protocol of the AW Suite Card IQ application, the software can reload saved tracking objects (Save State) incorrectly and display an incorrect vessel label over the restored images. This failure only occurs when users create a custom label in the vessel analysis protocol and include leading or trailing blank characters (space bar) in the custom label.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- GE Healthcare IITS sent Product Safety Notification letters dated 6/11/08 to all customers who have the AW Suite Software Versions 2.0 and 2.0.1, informing them that if they also have Card IQ Xpress Pro or Plus, the AW Suite Card IQ software can reload saved tracking objects (Save State) incorrectly and display an incorrect vessel label over the restored images. This failure only occurs when users create a custom label in the vessel analysis protocol and include leading or trailing blank characters (space bar) in the custom label. The users were instructed not to make any clinical decisions based on a Card IQ Save State when a custom label has been ordered, and not to include leading or trailing blank characters (space bar) in the new branch name when customizing a Card IQ vessel analysis protocol until the new software version can be installed. A software patch to permanently resolve the issue will be installed at no charge. Any questions were directed to the GE Customer Care Center at 800-437-1171.
- Quantity in commerce
- 207 units
- Distribution
- Nationwide and internationally to Australia, Belgium, Canada, Denmark, France, Germany, Great Britain, Hong Kong, Ireland, Israel, Italy, Japan, Malaysia, Malta, Portugal, Saudi Arabia, South Korea, Spain, Sweden, Switzerland and Turkey
- Product codes and lots
- Centricity AW Suite software version 2.0 and 2.0.1 with Card IQ Module