Z-2303-2008

Histochemical PAS (Periodic Acid Schiff) Reaction Set' Class I, IVD; Set of 4 bottles including Schiff...

EMD Millipore Corporation / initiated 2008-06-02 / Terminated / root cause: manufacturing process

EMD Chemicals Inc began a recall of Histochemical PAS (Periodic Acid Schiff) Reaction Set' Class I, IVD; Set of 4 bottles including Schiff Reagent 225 mL, Light Green SF Yellowish Stain 225 mL... on . The reason FDA records is "Two complaints that kit was not performing properly with positive controls". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2303-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
EMD Millipore Corporation
Product
Histochemical PAS (Periodic Acid Schiff) Reaction Set' Class I, IVD; Set of 4 bottles including Schiff Reagent 225 mL, Light Green SF Yellowish Stain 225 mL, Sodium Carbonate Solution 225 mL and Periodic Acid Solution 225 mL' EMD Chemicals, Inc. 480 S. Democrat Road, Gibbstown, NJ 08027 Each kit provides reagents sufficient for a maximum of 100 tests.
Product code
HZT Reagent, Schiff
Reason for recall
Two complaints that kit was not performing properly with positive controls.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Distributors were notified of the recall on June 4, 2008 by e-mail, followed by a letter. The letter dated June 4, 2008, describes the problem, instructs customers to discontinue use of the product and contact EMD Chemicals for product replacement. An Important Product Recall form was also included in each letter, asking that the form be completed and returned to EMD. Contact EMD Chemicals Technical Service Department at 1-800-222-0342.A
Quantity in commerce
29 kits
Distribution
Nationwide.
Product codes and lots
Item number 64945/93; Lot # 8055, Exp. August 31, 2009.