Update Instructions "TH003/08/S" applies to Siemens brand SIMIVIEW 3000, SIMVIEW NT, and MEV ASIM S systems...
Siemens Medical Solutions USA, Inc began a recall of Update Instructions "TH003/08/S" applies to Siemens brand SIMIVIEW 3000, SIMVIEW NT, and MEV ASIM S systems; Part Number: 5488460; Product is manufactured and... on . The reason FDA records is "The firm sent an Update Instruction, "TH003/08/S" for Motion Enable Switch/holder assembly. The switch holder allows the motion enable switch to be activated if inserted partially; thereby enabling inadvertent motion...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2196-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Update Instructions "TH003/08/S" applies to Siemens brand SIMIVIEW 3000, SIMVIEW NT, and MEV ASIM S systems; Part Number: 5488460; Product is manufactured and distributed by Siemens Medical Solutions, Oncology Care System, 4040 Nelson Ave, Concord CA The product is indicated for use in Radiation therapy simulation.
- Product code
- KPQ System, Simulation, Radiation Therapy
- Reason for recall
- The firm sent an Update Instruction, "TH003/08/S" for Motion Enable Switch/holder assembly. The switch holder allows the motion enable switch to be activated if inserted partially; thereby enabling inadvertent motion and serious injury may result.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On June 25, 2008 all consignees were notified via written letter and on-site visit by the firm's sales representative, informing them that the switch holder allows the motion enable switch to be activated if inserted partially; thereby enabling inadvertent motion and serious injury may result and providing instructions on the recall. A new software version will be provided. For assistance contact Siemens at 1-925-246-8407.
- Quantity in commerce
- 300 units
- Distribution
- Worldwide and Nationwide
- Product codes and lots
- All codes.