Z-0010-2009
Roche Diagnostics HDLC3, HDL-Cholesterol Plus 3rd Generation, COBAS INTEGRA, cobas c systems, Roche...
Roche Diagnostics GmbH / initiated 2008-06-11 / Terminated / root cause: labeling and UDI
Roche Diagnostics Corp. began a recall of Roche Diagnostics HDLC3, HDL-Cholesterol Plus 3rd Generation, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 04399803190. An in-vitro... on . The reason FDA records is "Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0010-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche Diagnostics HDLC3, HDL-Cholesterol Plus 3rd Generation, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 04399803190. An in-vitro diagnostic reagent system intended for use on COMAS Integra systems for the quantitative determination of HDL-cholesterol in serum and plasma.
- Product code
- LBS Ldl & Vldl Precipitation, Cholesterol Via Esterase-Oxidase, Hdl
- Reason for recall
- Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- Consignees were notified via an Urgent Medical Device correction letter dated 6/11/08 informing users of the problem and providing them with work-around instructions. The instructions suggests 1) inserting the cassette rack slowly; 2) using mainly reagent slots A, B, or C; 3) tilting the rack slightly clockwise so the barcode scanner does not read the barcode at an exact 90 degree angle; and 4) confirming the rack is properly seated on the track and displayed on the Cassette Status screen. If the cassette barcode does not register after following the Work-Around instructions, then customers are required to contact Diagnostics Technical Support.
- Quantity in commerce
- 1245
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot 69535801.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
