Z-0010-2009

Roche Diagnostics HDLC3, HDL-Cholesterol Plus 3rd Generation, COBAS INTEGRA, cobas c systems, Roche...

Roche Diagnostics GmbH / initiated 2008-06-11 / Terminated / root cause: labeling and UDI

Roche Diagnostics Corp. began a recall of Roche Diagnostics HDLC3, HDL-Cholesterol Plus 3rd Generation, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 04399803190. An in-vitro... on . The reason FDA records is "Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0010-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche Diagnostics HDLC3, HDL-Cholesterol Plus 3rd Generation, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 04399803190. An in-vitro diagnostic reagent system intended for use on COMAS Integra systems for the quantitative determination of HDL-cholesterol in serum and plasma.
Product code
LBS Ldl & Vldl Precipitation, Cholesterol Via Esterase-Oxidase, Hdl
Reason for recall
Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
Consignees were notified via an Urgent Medical Device correction letter dated 6/11/08 informing users of the problem and providing them with work-around instructions. The instructions suggests 1) inserting the cassette rack slowly; 2) using mainly reagent slots A, B, or C; 3) tilting the rack slightly clockwise so the barcode scanner does not read the barcode at an exact 90 degree angle; and 4) confirming the rack is properly seated on the track and displayed on the Cassette Status screen. If the cassette barcode does not register after following the Work-Around instructions, then customers are required to contact Diagnostics Technical Support.
Quantity in commerce
1245
Distribution
Nationwide Distribution
Product codes and lots
Lot 69535801.