Z-2197-2008

CABG PACK, Surgical Convenience Kit, Sterile, Reorder #:K12T-02449B, Merit Medical Systems, Inc., 1600 West...

Merit Medical Systems, Inc. / initiated 2008-03-14 / Terminated / root cause: packaging

Merit Medical Systems, Inc. began a recall of CABG PACK, Surgical Convenience Kit, Sterile, Reorder #:K12T-02449B, Merit Medical Systems, Inc., 1600 West Merit Parkway, South Jordan, Utah 84095. A... on . The reason FDA records is "Component of convenience kit may not be sterile". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2197-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
CABG PACK, Surgical Convenience Kit, Sterile, Reorder #:K12T-02449B, Merit Medical Systems, Inc., 1600 West Merit Parkway, South Jordan, Utah 84095. A combination of legally marketed medical devices placed into one container for the convenience of the user and used within the limits of the products' intended uses.
Product code
KDD Kit, Surgical Instrument, Disposable
Reason for recall
Component of convenience kit may not be sterile.
Root cause tag
packaging
FDA root cause
Packaging
Action
The sole direct account (distributor) and the sole end-user were notified by letter on 03/13-14/2008. Direct account was instructed to return any affected product still in inventory. End user was advised to discard the blades or re-sterilize them according to blade manufacturer's instructions, which were included. For assistance, contact Merit Medical at 1-801-208-4349.
Quantity in commerce
36 kits
Distribution
PA. No military, government, or foreign distribution.
Product codes and lots
Lot Number V583783