Sunrise Medical Quickie Groove Power Wheelchair with Transit option The product is a powered wheelchair for...
Sunrise Medical Inc began a recall of Sunrise Medical Quickie Groove Power Wheelchair with Transit option The product is a powered wheelchair for human use on . The reason FDA records is "Rear bolts holding the frame in place may shear and the front bolts may potentially come off the posts, resulting in partial or full seat detachment in a sudden stop or crash". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2339-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sunrise Medical, Inc.
- Product
- Sunrise Medical Quickie Groove Power Wheelchair with Transit option The product is a powered wheelchair for human use.
- Product code
- ITI Wheelchair, Powered
- Reason for recall
- Rear bolts holding the frame in place may shear and the front bolts may potentially come off the posts, resulting in partial or full seat detachment in a sudden stop or crash.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Initial phone contacts began on November 13, 2006. Letters were also mailed. The letter informed the customer of the problem and provided interim steps to take, until an transit upgrade kit is installed. Response cards will be tracked to measure response rate and assure receipt of notice. Contact Sunrise Medical at 1-800-456-8168 for assistance.
- Quantity in commerce
- 36 units
- Distribution
- Worldwide: US, Canada, New Zealand and Australia.
- Product codes and lots
- Multiple serial numbers, starting with GRV-100: 182, 181, 178, 176, 166, 161, 160, 151, 150, 115, 110, 109, 106, 100, 099, 098, 094, 082, 081, 079, 076, 073, 064, 058, 056, 054, 053, 050, 044, 043, 041, 040, 035, 033, 021 and 013