Z-2234-2008

StatSampler 200 uL, Li Heparin (micro blood collector) Ref: SS2H Packed in Catalog Numbers: SS2H (I vial)...

StatSpin, Inc., dba Iris Sample Processing / initiated 2008-06-20 / Terminated

StatSpin, Inc., dba Iris Sample Processing began a recall of StatSampler 200 uL, Li Heparin (micro blood collector) Ref: SS2H Packed in Catalog Numbers: SS2H (I vial), SS2H-06 (6 kits/case), CH2C-I0 (I0 vials/kit)... on . The reason FDA records is "EDTA inadvertently mixed with lithium heparin solution contained in capillary tubes". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2234-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
StatSpin, Inc., dba Iris Sample Processing
Product
StatSampler 200 uL, Li Heparin (micro blood collector) Ref: SS2H Packed in Catalog Numbers: SS2H (I vial), SS2H-06 (6 kits/case), CH2C-I0 (I0 vials/kit) Intended for collection and processing of capillary blood to prepare plasma/serum.
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
EDTA inadvertently mixed with lithium heparin solution contained in capillary tubes.
FDA root cause
Mixed-up of materials/components
Action
Iris Sample Processing notified customers by telephone followed by an Urgent: Product Recall letter dated 6/19/08. Users are requested to return inventory, to not use product from Lot Number 7296, and to repeat tests where the recalled tubes were used. The firm identified an additional lot : 8031 was effected and recalled the lot to two consignees.
Quantity in commerce
171 vials
Distribution
Worldwide Distribution --- USA including states of CA, CO, FL, KY, KS, GA, NC, NJ, OR, PA, and countries of Canada, Hong Kong, UK.
Product codes and lots
Lot number: 7296 Expiration Date: 12/09; Lot 8031 Expiration Date: 3/10