Z-2234-2008
StatSampler 200 uL, Li Heparin (micro blood collector) Ref: SS2H Packed in Catalog Numbers: SS2H (I vial)...
StatSpin, Inc., dba Iris Sample Processing / initiated 2008-06-20 / Terminated
StatSpin, Inc., dba Iris Sample Processing began a recall of StatSampler 200 uL, Li Heparin (micro blood collector) Ref: SS2H Packed in Catalog Numbers: SS2H (I vial), SS2H-06 (6 kits/case), CH2C-I0 (I0 vials/kit)... on . The reason FDA records is "EDTA inadvertently mixed with lithium heparin solution contained in capillary tubes". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2234-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- StatSpin, Inc., dba Iris Sample Processing
- Product
- StatSampler 200 uL, Li Heparin (micro blood collector) Ref: SS2H Packed in Catalog Numbers: SS2H (I vial), SS2H-06 (6 kits/case), CH2C-I0 (I0 vials/kit) Intended for collection and processing of capillary blood to prepare plasma/serum.
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- EDTA inadvertently mixed with lithium heparin solution contained in capillary tubes.
- FDA root cause
- Mixed-up of materials/components
- Action
- Iris Sample Processing notified customers by telephone followed by an Urgent: Product Recall letter dated 6/19/08. Users are requested to return inventory, to not use product from Lot Number 7296, and to repeat tests where the recalled tubes were used. The firm identified an additional lot : 8031 was effected and recalled the lot to two consignees.
- Quantity in commerce
- 171 vials
- Distribution
- Worldwide Distribution --- USA including states of CA, CO, FL, KY, KS, GA, NC, NJ, OR, PA, and countries of Canada, Hong Kong, UK.
- Product codes and lots
- Lot number: 7296 Expiration Date: 12/09; Lot 8031 Expiration Date: 3/10