Z-2226-2008
Gyrus ACMI 10 Fr. Dual Lumen Catheter Reference Number: 608 BX (Box of 5) and 608 (single unit) This product...
Gyrus ACMI, Inc. / initiated 2008-06-25 / Terminated / root cause: manufacturing process
Gyrus ACMI Corporation began a recall of Gyrus ACMI 10 Fr. Dual Lumen Catheter Reference Number: 608 BX (Box of 5) and 608 (single unit) This product has multiple uses, including ureteral dilation... on . The reason FDA records is "Particulate matter in the catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2226-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Gyrus ACMI, Inc.
- Product
- Gyrus ACMI 10 Fr. Dual Lumen Catheter Reference Number: 608 BX (Box of 5) and 608 (single unit) This product has multiple uses, including ureteral dilation, anesthetic injection, stone displacement, contrast injection, and safety wire / guidewire placement. The "ACMI" Dual Lumen Catheter is a radiopaque ureteral catheter.
- Product code
- EYB Catheter, Ureteral, Gastro-Urology
- Reason for recall
- Particulate matter in the catheter
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Gyrus initiated a recall by letter dated 6/ 25 /08. An updated letter issued July 9, 2008 which supersedes the previous letter. The July 9th letter updates the lot numbers affected in the recall. A reply form request was sent to consignees informing them of the problem and asking them to obtain a RMA and return the product. They are also requested to FAX back a reply form so that the response can be tracked. For assistance, please contact Gyrus ACMI at 1-888-524-7266.
- Quantity in commerce
- 6820 units
- Distribution
- Nationwide Foreign: Canada, UK, Netherlands, Spain, Italy, Sweden, France, Portugal, Ireland, , Brazil, Australia, South Africa, and Turkey
- Product codes and lots
- Lot Numbers: 7124026,7131020,7138014, 7152018,7158015,7169005,7201018,721 1016,72I404l, 7227021,7232046,7253048, 7256034,7263052,7269089,7276035,7277024,7284042,7291034,7296052,7303068, 7306089,7312076,7324062,7337052,7344047,7347078,7354102,7361043,8010043, 8017047,8025006, 8031030, 8036057, 8045050, 8052064, 8059055, 8066067, and 8073030