Z-1396-2009

Access Myoglobin Reagents Kits, Part Number: 973243. The Access Myoglobin assay is a paramagnetic particle...

Siemens Healthcare Diagnostics Inc. / initiated 2007-09-27 / Terminated / root cause: labeling and UDI

Beckman Coulter Inc began a recall of Access Myoglobin Reagents Kits, Part Number: 973243. The Access Myoglobin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative... on . The reason FDA records is "Beckman Coulter initiated the field action as a result of recent internal testing following the CLSI EP-5A - Evaluation of Precision Performance of Clinical Chemistry Devices protocol, it was determined that the total...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1396-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Myoglobin Reagents Kits, Part Number: 973243. The Access Myoglobin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Myoglobin levels in human serum and plasma using the Access Immunoassay Systems.
Product code
DDR Myoglobin, Antigen, Antiserum, Control
Reason for recall
Beckman Coulter initiated the field action as a result of recent internal testing following the CLSI EP-5A - Evaluation of Precision Performance of Clinical Chemistry Devices protocol, it was determined that the total imprecision for the Access Myoglobin Reagent is greater than or equal to 10% across the expected physiological range rather than the stated claim less than 8%. Based on internal tes
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
An Urgent: Product Corrective Action (PCA) letter was mailed via US Postal Service on October 08, 2007 to the customers that received Access Myoglobin Reagent kits informing them about the issue. The letter also informed customers that the product insert in the Access Assay Manual will be revised with the new range to reflect the new imprecision claim. Customers were instructed to: (1) Share this information with their laboratory staff and retain the letter for their Quality System documentation, (2) Complete and return the enclosed fax-back response form within 10 days so that Beckman Coulter, Inc. can be assured that the customer received the notification, and (3) Contact Customer Technical Support at 1-800-854-3633 or your local Beckman Coulter, Inc. representative for questions or additional information regarding the notification.
Quantity in commerce
20,727
Distribution
Nationwide and Canada.
Product codes and lots
All Lot Numbers.