Siemens brand Ultrasound Transducer 17L5HD used with ACUSON Sequoia Ultrasound System; Model Number(s)...
Siemens Medical Solutions USA, Inc. began a recall of Siemens brand Ultrasound Transducer 17L5HD used with ACUSON Sequoia Ultrasound System; Model Number(s): 10035724, used with System Models: 08245875, 08246951... on . The reason FDA records is "The product assembly can cause a failure of the transducer which results in a double and/or overlapped image which may ultimately result in misleading or false information, inability to accurately diagnose, incorrect...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2230-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens brand Ultrasound Transducer 17L5HD used with ACUSON Sequoia Ultrasound System; Model Number(s): 10035724, used with System Models: 08245875, 08246951, 08255412, 08255413, 08267697, 08269627, 10038241, 10038242, 10040724, 10040725, 10041008; Product is manufactured and distributed by Siemens Medical Solutions USA, Inc. Ultrasound Division, Mountain View, CA The 17L5 transducer is intended for breast and small body parts imaging.
- Product code
- IYO System, Imaging, Pulsed Echo, Ultrasonic
- Reason for recall
- The product assembly can cause a failure of the transducer which results in a double and/or overlapped image which may ultimately result in misleading or false information, inability to accurately diagnose, incorrect positioning/locating/insertion of the biopsy needle and/or an interruption during the biopsy procedure.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 6/6/08, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the affected product be returned. Affected transducers will be replaced. For assistance, contact Siemens Medical Solutions at 1-650-694-5993.
- Quantity in commerce
- 388 units
- Distribution
- Worldwide Distribution: USA, AUSTRALIA, AUSTRIA, CANADA, CHINA, DENMARK, FRANCE, GERMANY, INDIA, ITALY, JAPAN, KOREA, MALAYSIA, MEXICO, NORWAY, PORTUGAL, RUSSIAN FEDERATION, SINGAPORE, SWEDEN, TAIWAN, and THAILAND.
- Product codes and lots
- Serial numbers beginning with 735 and lower.