Z-2229-2008

Naviscan PET Systems, Inc ("Naviscan") PEM Flex Solo II PET Scanner. The PEM Flex Solo II images small body...

CMR Naviscan Corporation / initiated 2008-06-19 / Terminated / root cause: design

Naviscan PET Systems began a recall of Naviscan PET Systems, Inc ("Naviscan") PEM Flex Solo II PET Scanner. The PEM Flex Solo II images small body parts with optimized PET technology on . The reason FDA records is "This field correction is being initiated as a result of the firm's investigation into a report from a user indicating that the motorized compression exceeded 25 pounds of compression force during the pre-scan...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2229-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CMR Naviscan Corporation
Product
Naviscan PET Systems, Inc ("Naviscan") PEM Flex Solo II PET Scanner. The PEM Flex Solo II images small body parts with optimized PET technology.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
This field correction is being initiated as a result of the firm's investigation into a report from a user indicating that the motorized compression exceeded 25 pounds of compression force during the pre-scan positioning of the patient. The investigation identified that a secondary fault protection did not perform as expected.
Root cause tag
design
FDA root cause
Device Design
Action
An Urgent Medical Device Field Correction Notice was sent to all PEM Flex Solo II customers about this issue via Federal Express Priority Service on June 19, 2008. Customers were informed of the: (1) Issue: Potential for motorized compression to exceed 25 pounds of compression force if continued beyond the primary safety limit, This could lead, to patient discomfort or injury. (2) Advisory: Naviscan recommends that patient breast positioning be completed by ceasing motorized-compression when the primary safety limit of 15 pounds pre-compression force is reached. This is indicated by the automatic stopping of the paddle movement as controlled by the primary safety system. "Fine-tuning" positioning, should be completed by using the manual adjustment knobs. (3) Correction: A Naviscan Field Service Engineer will contact each customer to schedule a field correction visit. If customers have any questions they were instructed to contact Naviscan Customer Support at 1-888-628-4159.
Quantity in commerce
13
Distribution
Nationwide to CA, FL, MA, MO, NY, OK, PA, TX, and WA
Product codes and lots
Serial Numbers: P002-B-070316-01, P002-B-070423-01, P002-B-070531-01, P002-B-070629-01, P002-B-070712-0l, P002-B-070718-01, P002-B-070910-01, P002-B-070911-01, P002-B-071015-01, P002-B-080107-01, P002-B-080108-01, P002-B-080207-01, and P002-B-080307-01