Reflex Revision Screwdriver Inner Shaft is a orthopedic manual surgical instrument that is a nonpowered, hand...

Howmedica Osteonics Corp / initiated 2008-06-05 / Terminated / root cause: manufacturing process

Stryker Spine began a recall of Reflex Revision Screwdriver Inner Shaft is a orthopedic manual surgical instrument that is a nonpowered, hand held device intended for medical purposes to... on . The reason FDA records is "The Inner Shaft does not extend far enough from the tip of the Revision Screwdriver to fully engage the bone screw". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2238-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Reflex Revision Screwdriver Inner Shaft is a orthopedic manual surgical instrument that is a nonpowered, hand held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery. Ref Number 48500900;Non Sterile
Product code
HXX Screwdriver
Reason for recall
The Inner Shaft does not extend far enough from the tip of the Revision Screwdriver to fully engage the bone screw.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
URGENT PRODUCT RECALL letters were sent to all branches/agencies who received the product on June 5, 2008. If you have questions, contact Michelle Klein at 201-760-8256.
Quantity in commerce
35
Distribution
The product was distributed to 13 Stryker branches/agencies.
Product codes and lots
Lot codes: 26719, 047377, 047570, 065167, 06A167, and 072689.