Z-2238-2008
Reflex Revision Screwdriver Inner Shaft is a orthopedic manual surgical instrument that is a nonpowered, hand...
Howmedica Osteonics Corp / initiated 2008-06-05 / Terminated / root cause: manufacturing process
Stryker Spine began a recall of Reflex Revision Screwdriver Inner Shaft is a orthopedic manual surgical instrument that is a nonpowered, hand held device intended for medical purposes to... on . The reason FDA records is "The Inner Shaft does not extend far enough from the tip of the Revision Screwdriver to fully engage the bone screw". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2238-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Reflex Revision Screwdriver Inner Shaft is a orthopedic manual surgical instrument that is a nonpowered, hand held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery. Ref Number 48500900;Non Sterile
- Product code
- HXX Screwdriver
- Reason for recall
- The Inner Shaft does not extend far enough from the tip of the Revision Screwdriver to fully engage the bone screw.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- URGENT PRODUCT RECALL letters were sent to all branches/agencies who received the product on June 5, 2008. If you have questions, contact Michelle Klein at 201-760-8256.
- Quantity in commerce
- 35
- Distribution
- The product was distributed to 13 Stryker branches/agencies.
- Product codes and lots
- Lot codes: 26719, 047377, 047570, 065167, 06A167, and 072689.