Z-2240-2008

NobelReplace Tapered Groovy RP 4.3x10mm. These products are root-form endosseous implants intended to be...

Nobel Biocare USA LLC / initiated 2007-11-02 / Terminated / root cause: labeling and UDI

Nobel Biocare USA LLC began a recall of NobelReplace Tapered Groovy RP 4.3x10mm. These products are root-form endosseous implants intended to be surgically placed in the bone of the upper or lower... on . The reason FDA records is "This recall was initiated after it was discovered that the NobelReplace Tapered Groovy RP 4.3x10mm, REF 32216 Lot#403504 has an incorrect cap label. The affected lot has a cap label identifying the implant as "4.3x8"...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2240-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Nobel Biocare USA LLC
Product
NobelReplace Tapered Groovy RP 4.3x10mm. These products are root-form endosseous implants intended to be surgically placed in the bone of the upper or lower jaw arches to support for prosthetic devices.
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
This recall was initiated after it was discovered that the NobelReplace Tapered Groovy RP 4.3x10mm, REF 32216 Lot#403504 has an incorrect cap label. The affected lot has a cap label identifying the implant as "4.3x8". Since this cap label may be the only means used for identifying the implant prior to surgery, it is possible that a user would try to use the 10mm implant as an 8mm implant.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
The recall was initiated on 11/02/2007 with telephone calls to the affected customers. A "Urgent Medical Device Recall" letter with attached Customer Response form was then sent via US Postal Service beginning on 11/02/2007. The letter instructed customers to please immediately remove and return REF 32216 Lot #403504, NobelReplace Tapered Groovy RP 4.3x10mm. The firm contracted UPS courier service to pick up the recalled products. Customers were also instructed if the have further distributed the affected product to please immediately notify those customer to cease using this particular lot and forward the customer this notice. The firm indicated in the recall notice that they are in the process of sending out replacement products. Contact Kam Leung at 714-282-4800 if you have questions.
Quantity in commerce
197
Distribution
AR, AZ, CA, CO, FL, GA, IA, IL, MA, ME, MI, MO, NC, NC, NJ, NY, OK, PA, TX, UT, VA, WA
Product codes and lots
Ref 32216 lot number: 403504