Z-2242-2008

Cannon Catheter" II, Long-Term Access. Catheter Length, Cuff to Tip: 19cm, CS-15242, Intended for single use...

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2008-06-10 / Terminated / root cause: design

Arrow International Inc began a recall of Cannon Catheter" II, Long-Term Access. Catheter Length, Cuff to Tip: 19cm, CS-15242, Intended for single use only, Sterile, Rx only, Arrow International, Inc... on . The reason FDA records is "The tips may not have been adequately welded to the catheter body". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2242-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Cannon Catheter" II, Long-Term Access. Catheter Length, Cuff to Tip: 19cm, CS-15242, Intended for single use only, Sterile, Rx only, Arrow International, Inc., 2400 Bernville Road, Reading, PA 19605, USA The product is used for long term hemodialysis vascular access.
Product code
MSD Catheter, Hemodialysis, Implanted
Reason for recall
The tips may not have been adequately welded to the catheter body.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were notified by letter on/about 06/10/2008. They were instructed to: 1. Cease use and distribution and quarantine all affected product immediately. 2. Contact Arrows Customer Service Department for a Return Authorization Number, enter the number on an enclosed Recall Acknowledgement & Stock Status Form and 3. Immediately fax back the form to Arrow International. All affected product is to be returned to Arrow International. A letter containing the recall information was enclosed for physicians who have implanted catheters. Contact Arrow's Customer Service at 1-800-523-8446 for assistance.
Quantity in commerce
79248 units (total for all catheters listed)
Distribution
U.S. Nationwide, Albania, Argentina, Australia, Belgium, Brazil, Canada , Chile, Colombia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Guatemala, Hungary, India, Italy, Latvia, Liechtenstein, Lithuania, Mexico, Netherlands, New Zealand, Pakistan, Poland, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, Tanzania, Turkey, UK and the Ukraine
Product codes and lots
Lot #RL7065076, RL7054704, RL7034338, RL6123772, RL6113608, RL6103456, and RL6062695.