Z-2286-2008

Piccolo Chemistry Analyzer software version 6.121 with Comprehensive Metabolic Lot #8111AA5 , Model Number...

Abaxis Inc / initiated 2008-04-21 / Terminated / root cause: manufacturing process

Abaxis Inc began a recall of Piccolo Chemistry Analyzer software version 6.121 with Comprehensive Metabolic Lot #8111AA5 , Model Number 400-1028 in polyfoil pouch. Piccolo Comprehensive... on . The reason FDA records is "The product will recover tCO2 at approximately 5-7 units high relative to some controls and patient samples". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2286-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abaxis Inc
Product
Piccolo Chemistry Analyzer software version 6.121 with Comprehensive Metabolic Lot #8111AA5 , Model Number 400-1028 in polyfoil pouch. Piccolo Comprehensive Metabolic Panel is a single-use self-contained reagent disk, intended for use with the Piccolo and PiccoloXpress chemistry analyzers, for in vitro qualitative determination.
Product code
JLB Atp And Creatine Kinase (Enzymatic), Creatine
Reason for recall
The product will recover tCO2 at approximately 5-7 units high relative to some controls and patient samples
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm issued a notification letter to its consignees on April 21, 2008. The firm intends to follow up with phone calls and provide a software upgrade for the products. Contact Abaxis Customer Service at 1-510-675-6500 for assistance.
Quantity in commerce
893 kits
Distribution
Product was distributed Worldwide.
Product codes and lots
software version 6.121 with Comprrehensive Metabolic Lot #8111AA5, expiration date 2009-03-10