Z-2286-2008
Piccolo Chemistry Analyzer software version 6.121 with Comprehensive Metabolic Lot #8111AA5 , Model Number...
Abaxis Inc / initiated 2008-04-21 / Terminated / root cause: manufacturing process
Abaxis Inc began a recall of Piccolo Chemistry Analyzer software version 6.121 with Comprehensive Metabolic Lot #8111AA5 , Model Number 400-1028 in polyfoil pouch. Piccolo Comprehensive... on . The reason FDA records is "The product will recover tCO2 at approximately 5-7 units high relative to some controls and patient samples". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2286-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abaxis Inc
- Product
- Piccolo Chemistry Analyzer software version 6.121 with Comprehensive Metabolic Lot #8111AA5 , Model Number 400-1028 in polyfoil pouch. Piccolo Comprehensive Metabolic Panel is a single-use self-contained reagent disk, intended for use with the Piccolo and PiccoloXpress chemistry analyzers, for in vitro qualitative determination.
- Product code
- JLB Atp And Creatine Kinase (Enzymatic), Creatine
- Reason for recall
- The product will recover tCO2 at approximately 5-7 units high relative to some controls and patient samples
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm issued a notification letter to its consignees on April 21, 2008. The firm intends to follow up with phone calls and provide a software upgrade for the products. Contact Abaxis Customer Service at 1-510-675-6500 for assistance.
- Quantity in commerce
- 893 kits
- Distribution
- Product was distributed Worldwide.
- Product codes and lots
- software version 6.121 with Comprrehensive Metabolic Lot #8111AA5, expiration date 2009-03-10