Z-2296-2008

Zimmer Patient Helper Reinforced Overhead Bar for Hill-Rom beds with removable headboards; Zimmer Inc...

Zimmer GmbH / initiated 2008-07-07 / Terminated / root cause: design

Zimmer Inc. began a recall of Zimmer Patient Helper Reinforced Overhead Bar for Hill-Rom beds with removable headboards; Zimmer Inc., Warsaw, IN; Prod. Nos. 00-0965-000-00 and... on . The reason FDA records is "If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2296-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Patient Helper Reinforced Overhead Bar for Hill-Rom beds with removable headboards; Zimmer Inc., Warsaw, IN; Prod. Nos. 00-0965-000-00 and 00-0965-001-00. The product is a L shaped bar that is used with hospital beds set up for traction. The bar allows mounting of a trapeze to help the patient pull themselves up when in the hospital bed.
Product code
IKX Aid, Transfer
Reason for recall
If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were notified via recall letter dated 7/7/08 to discontinue use of these devices until the firm visits and replaces the devices. Contact Zimmer at 1-800-321-5533 for assistance.
Quantity in commerce
645 of all systems
Distribution
Nationwide and Canada.
Product codes and lots
All units shipped between 7/11/07 and 5/19/08.