Z-2301-2008

Zimmer NEXGEN Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor (a pounding instrument...

Zimmer GmbH / initiated 2008-06-30 / Terminated / root cause: design

Zimmer Inc. began a recall of Zimmer NEXGEN Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor (a pounding instrument used during knee surgery), nonsterile, Zimmer... on . The reason FDA records is "The instrument may fracture during use, resulting in metal fragments being left in the patient post-surgery, which could cause implant failure". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2301-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer NEXGEN Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor (a pounding instrument used during knee surgery), nonsterile, Zimmer, Warsaw, Indiana; REF 5951-90..
Product code
HWA Impactor
Reason for recall
The instrument may fracture during use, resulting in metal fragments being left in the patient post-surgery, which could cause implant failure.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were notified via letter dated 6/30/08. The letter described the issue, provided instructions for the return of the product and asked customers to complete a reply back form. For assistance, contact Zimmer at 1-574-372-4487.
Quantity in commerce
1439
Distribution
Nationwide, Australia, China, England, France, Germany, India, Italy, Japan, Korea, Singapore and Taiwan.
Product codes and lots
All units.