NeoMedix Trabectome I/A Console, Model Number 550014, surgical aspirator
Neomedix Corp began a recall of NeoMedix Trabectome I/A Console, Model Number 550014, surgical aspirator on . The reason FDA records is "Failure of irrigation flow: The pinch valve in a few pumps may not consistently open to allow irrigation flow. In over 450 surgeries, the firm confirmed one instance of a valve that did not open. The electrically...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0071-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Neomedix Corp.
- Product
- NeoMedix Trabectome I/A Console, Model Number 550014, surgical aspirator
- Product code
- BTA Pump, Portable, Aspiration (Manual Or Powered)
- Reason for recall
- Failure of irrigation flow: The pinch valve in a few pumps may not consistently open to allow irrigation flow. In over 450 surgeries, the firm confirmed one instance of a valve that did not open. The electrically activated pinch valve in the NeoMedix Trabectome I/A pump activates irrigation flow to the eye during Trabectome surgery.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On May 31, 2007 NeoMedix mailed a Customer letter to all consignees informing them of the hazard. The letter also indicated to the customers they NeoMedix has developed a minor modification/upgrade that is designed to assure that the pinch valve opens at the appropriate time. The customers were instructed this modification requires return of their existing pump for modification and testing at the firm's facility before its return. In the event that surgeries have already been scheduled, NeoMedix will provide a loaner unit until the suspect unit can be modified and returned to you. The letter also indicated the cost of the modification and/or use of a loaner unit will be entirely at NeoMedix expense. You will be contacted to establish the details of implementing this modification.
- Quantity in commerce
- 7
- Distribution
- Nationwide to: CA, HI, IA, NY & PA
- Product codes and lots
- Model Number 550014; Lot Number 061205-1; Serial Numbers: 10117, 10119, 10120, 10122, 10124, 100129, and 10133.