Z-2306-2008
Corflo Anti-I.V. NG Tube for Pediatric and Neonatal Use; a sterile clear polyurethane with orange radiopaque...
Medtronic, Inc. / initiated 2008-06-25 / Terminated
Cardinal Health began a recall of Corflo Anti-I.V. NG Tube for Pediatric and Neonatal Use; a sterile clear polyurethane with orange radiopaque stripe Enteral Feeding Tube without Stylet, 5 Fr... on . The reason FDA records is "The male Luer adapter may be the wrong component therefore the feeding set would not be able to be connected to the tube nor would the cap fit securely". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2306-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Corflo Anti-I.V. NG Tube for Pediatric and Neonatal Use; a sterile clear polyurethane with orange radiopaque stripe Enteral Feeding Tube without Stylet, 5 Fr, 22" long, non-weighted, DEHP and latex free; individually packaged, 10 tubes per carton; Mfg. by Viasys MedSystems, Wheeling, IL 60090 Product is intended for use in those patients who require intermittent or continuous tube feedings via the nasogastric or nasoenteric route.
- Product code
- KNT Tubes, Gastrointestinal (And Accessories)
- Reason for recall
- The male Luer adapter may be the wrong component therefore the feeding set would not be able to be connected to the tube nor would the cap fit securely.
- FDA root cause
- Mixed-up of materials/components
- Action
- Cardinal Health notified the accounts who ordered the affected feeding tubes via letter dated 6/25/08, informing them that the feeding tube may have the wrong male Luer adapter component, which can cause leakage around the tube outside of the body. Upon review of the letter, the accounts were requested to inspect their inventory and immediately destroy any affected product found and to telephone Cardinal Health/VIASYS MedSystems Technical Service at 1-800-323-6305 for arrange for credit. Dealers were requested to notify their customers. Any questions were directed to Cardinal Health/VIASYS MedSystems at 800-403-3400, ext. 3473. The accounts were also requested to complete and fax back the enclosed acknowledgement form indicating if they have any remaining inventory of the recalled scissors or not.
- Quantity in commerce
- 807 tubes
- Distribution
- California, Illinois, Kentucky and Missouri
- Product codes and lots
- catalog/reorder number 20-1225AIV, lot 32601