The St. Jude Medical Livewire TC" Ablation Catheter, 8mm Tip w/Thermocouple & Thermistor REF 402196, Length...
St. Jude Medical Cardiovascular Division began a recall of The St. Jude Medical Livewire TC" Ablation Catheter, 8mm Tip w/Thermocouple & Thermistor REF 402196, Length 115 cm, Lot 2056960, Manufacturer St. Jude Medical... on . The reason FDA records is "St. Jude Medical has initiated a recall of one lot of Livewire TC" Ablation Catheter. The product was mis-labelled as having an 8 mm tip instead of 4 mm tip. This could potentially result in elevated temperatures and...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2313-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- The St. Jude Medical Livewire TC" Ablation Catheter, 8mm Tip w/Thermocouple & Thermistor REF 402196, Length 115 cm, Lot 2056960, Manufacturer St. Jude Medical 14901 DeVeau Place, Minnetonka, MN 55345-2126 USA, Sterile EO The St. Jude Medical Livewire TC" Ablation Catheter is a flexible electrode catheter constructed of a radiopaque polyurethane insulation/shaft and incorporated platinum electrode. The distal tip portion may be deflected by a remote control handle located a t the proximal end of t the catheter.
- Product code
- LPB Cardiac Ablation Percutaneous Catheter
- Reason for recall
- St. Jude Medical has initiated a recall of one lot of Livewire TC" Ablation Catheter. The product was mis-labelled as having an 8 mm tip instead of 4 mm tip. This could potentially result in elevated temperatures and formation of coagulum at the tip electrode. If this coagulum were to detach from the product, the worst case this could potentially result in embolization.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Consignees were sent a "St Jude Medical Urgent Product Recall" letter addressed to "Dear Doctor" and a "Doctor's Acknowledgement Form" . The letter described the product, problem and the possibility of a health hazard. They also requested discontinue of product and gave instructions for return of product. Contact St. Jude Medical at 952-933-4700 if you have questions.
- Quantity in commerce
- 39
- Distribution
- Canada and Belgium
- Product codes and lots
- Lot # 2056960