Z-2321-2008

Medline RF Detect Sterile X-Ray Detectable USP Type VII Gauze; 8" x 4", 12-ply, 100% cotton sponges, 10...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2008-06-13 / Terminated / root cause: design

Medline Industries Inc began a recall of Medline RF Detect Sterile X-Ray Detectable USP Type VII Gauze; 8" x 4", 12-ply, 100% cotton sponges, 10 sponges per pack, 80 packs per case; Manufactured for... on . The reason FDA records is "There is the potential for the RF ID tag to separate from the retaining pouch". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2321-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline RF Detect Sterile X-Ray Detectable USP Type VII Gauze; 8" x 4", 12-ply, 100% cotton sponges, 10 sponges per pack, 80 packs per case; Manufactured for: Medline Industries, Inc., Mundelein, IL 60060 USA, Made in China; reorder: NON21432RF
Product code
GDY Gauze/Sponge, Internal, X-Ray Detectable
Reason for recall
There is the potential for the RF ID tag to separate from the retaining pouch.
Root cause tag
design
FDA root cause
Device Design
Action
Initially a recall letter dated 6/13/08 was sent to the direct accounts identifying 5 lots of affected lots with two reorder numbers involved, explaining that there was a lack of integrity of the retaining pouch material for the RFID gauze, and that there was a potential for the RFID tag to separate from the retaining pouch during use. It was subsequently discovered that there was only one reorder number affected by the problem and that an additional lot had been gamma sterilized. Therefore a second letter dated 6/19/08, listing the correct reorder number and the six affected lots, was sent via FedEx overnight on the same date. The accounts were requested to remove RFID gauzes with the listed lot numbers from all stock rooms, materials storage, and OR's, and return it to Medline for replacement. The accounts were requested to complete the enclosed response sheet and fax it back to Medline at 847-643-4482, indicating the amount and lot numbers of the gauzes being returned. Any questions were directed to 866-359-1704, between 8:00 AM and 5:00 PM CST. Distributors were requested to notify their accounts of the recall. If you have questions about this recall, contact Kathy Dunne at 847-970-4678 or at [email protected].
Quantity in commerce
295 cases
Distribution
California, Indiana, Missouri, Ohio, Oklahoma, Pennsylvania, Texas and Washington
Product codes and lots
reorder: NON21432RF, lots CM130002052, CM130002062, CM130002075, CM130002068, CM130002079, and CM130002110