Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2322-2008

Medtronic iCon Patient Programmer, model 3037. The Patient programmer is a hand held, battery operated...

Medtronic Neuromodulation / initiated 2008-05-12 / Terminated / root cause: manufacturing process

Medtronic Neuromodulation began a recall of Medtronic iCon Patient Programmer, model 3037. The Patient programmer is a hand held, battery operated, microcontroller based device for use by patients to... on . The reason FDA records is "Medtronic is retrieving four (4) patient programmers that were not properly loaded with application software. The application software is needed for a patient programmer to synchronize and bond with a neurostimulation...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2322-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Medtronic iCon Patient Programmer, model 3037. The Patient programmer is a hand held, battery operated, microcontroller based device for use by patients to control and monitor external and implantable devices. The minimum amount of control involves the ability to turn therapy ON or OFF; maximal control allows the patient to select from different therapies and adjust the parameters with in the set therapy limits.
Product code
EZW Stimulator, Electrical, Implantable, For Incontinence
Reason for recall
Medtronic is retrieving four (4) patient programmers that were not properly loaded with application software. The application software is needed for a patient programmer to synchronize and bond with a neurostimulation device. Without this functionality a patient programmer is not useable and can not communicate with a neurostimulation device.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
A letter was sent to each account telling them of the issue and to let them know that Medtronic is retrieving the affected devices. Medtronic Reps were to visit affected accounts to notify them of the recall, locate the affected device and return it. Rep is to complete the reply card and fax to Medtronic Neuromodulation. If you have questions, contact your Medtronic representative or Medtronic Neuromodulation Technical Services at 1-800-707-0933.
Quantity in commerce
2
Distribution
GA, MO, NC, TX no OUS distribution
Product codes and lots
serial numbers: NJD058707N, NJD062024N, NJD063259N, and NJD063324N