Z-2389-2008

NicoletOne photic adapter cable 085-463700

Medtronic, Inc. / initiated 2008-06-05 / Terminated / root cause: manufacturing process

Cardinal Health NeuroCare Division began a recall of NicoletOne photic adapter cable 085-463700 on . The reason FDA records is "Cardinal Health NeuroCare is implementing a voluntary field correction to replace a quantity of photic adapter cables, part number 085-463700, used to connect the NicLED Photic Strobe option, part number 698-625500...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2389-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
NicoletOne photic adapter cable 085-463700
Product code
GWQ Full-Montage Standard Electroencephalograph
Reason for recall
Cardinal Health NeuroCare is implementing a voluntary field correction to replace a quantity of photic adapter cables, part number 085-463700, used to connect the NicLED Photic Strobe option, part number 698-625500, with NicoletOne EEG system. The photic adapter cable has a wiring error which delivers a reduced voltage and results in a decreased intensity of the NicLED photic. This reduced intensi
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
A "Cardinal Health Urgent: Medical Device Correction" letter was sent to consignees. The letter described the product and the problem. The consignees were asked to confirm completion of the cable exchange by returning an information card. If you have questions, contact Cardinal Health (formerly VIASYS) NeuroCare Technical Support Group at 1-800-356-0007 (US), 608-273-5000 (OUS) or email [email protected].
Quantity in commerce
92
Distribution
CA FL GA ID IN LA MD ME MI MO NC NY PA TN TX WA WI China Denmark Indonesia Japan Korea Mexico Netherlands Russia Saudi Arabia Slovenia
Product codes and lots
085-463700