Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2381-2008

Medtronic Intrathecal Catheter, model 8731SC. 38.1 cm spinal segment with length markers and guide wire, 66...

Medtronic Neuromodulation / initiated 2008-06-26 / Terminated / root cause: design

Medtronic Neuromodulation began a recall of Medtronic Intrathecal Catheter, model 8731SC. 38.1 cm spinal segment with length markers and guide wire, 66 cm pump segment, separate connector pin, 15T gauge... on . The reason FDA records is "Disconnection of the Medtronic sutureless connector (SC) catheters from the catheter port on the pump, or occlusion between the sutureless pump connector and the catheter port on the pump". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2381-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Intrathecal Catheter, model 8731SC. 38.1 cm spinal segment with length markers and guide wire, 66 cm pump segment, separate connector pin, 15T gauge introducer needles (2) (11.4 cm and 9.3 cm), Transparent strain-relief sleeves (2), Opaque Strain-relief sleeves (2), V-wing Anchors (2).
Product code
LKK Pump, Infusion, Implanted, Programmable
Reason for recall
Disconnection of the Medtronic sutureless connector (SC) catheters from the catheter port on the pump, or occlusion between the sutureless pump connector and the catheter port on the pump.
Root cause tag
design
FDA root cause
Component design/selection
Action
Medtronic sent A Safety Alert dated June 2008, on June 27, 2008, to healthcare professionals, which described the issues of occlusion and disconnect. The letter identifies the affected catheter and revision kit model numbers and the associated implantable infusion pumps. The issues do not involve Medtronic MiniMed insulin pump. The letter refers the HCP to recommendations for Implant Techniques and Recommendations for Patency Verification that are included with the letter. The letter also includes recommendations for managing patients with implanted SC catheters.
Quantity in commerce
21,674 total
Distribution
Worldwide Distribution: USA, Puerto Rico, Algeria, Aruba, Austria, Australia, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malta, Netherlands, Netherlands Antilles, Norway, Poland, Portugal, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom.
Product codes and lots
All SC Catheters