Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: Catheter interface with attached...
Medtronic Neuromodulation began a recall of Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: Catheter interface with attached sutureless pump connector, catheter, connector pin... on . The reason FDA records is "Disconnection of the Medtronic sutureless connector (SC) catheters from the catheter port on the pump, or occlusion between the sutureless pump connector and the catheter port on the pump". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2382-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Neuromodulation
- Product
- Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: Catheter interface with attached sutureless pump connector, catheter, connector pin, and strain-relief sleeve, Strain-relief Sleeves (2).
- Product code
- LKK Pump, Infusion, Implanted, Programmable
- Reason for recall
- Disconnection of the Medtronic sutureless connector (SC) catheters from the catheter port on the pump, or occlusion between the sutureless pump connector and the catheter port on the pump.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Medtronic sent A Safety Alert dated June 2008, on June 27, 2008, to healthcare professionals, which described the issues of occlusion and disconnect. The letter identifies the affected catheter and revision kit model numbers and the associated implantable infusion pumps. The issues do not involve Medtronic MiniMed insulin pump. The letter refers the HCP to recommendations for Implant Techniques and Recommendations for Patency Verification that are included with the letter. The letter also includes recommendations for managing patients with implanted SC catheters.
- Quantity in commerce
- 21,674 total
- Distribution
- Worldwide Distribution: USA, Puerto Rico, Algeria, Aruba, Austria, Australia, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malta, Netherlands, Netherlands Antilles, Norway, Poland, Portugal, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom.
- Product codes and lots
- All SC Catheter Revision Kits