Z-2326-2008

Evita 4 Critical Care Ventilator. Continuous (Respirator)

Draeger, Inc. / initiated 2008-07-02 / Terminated / root cause: manufacturing process

Draeger Medical, Inc. began a recall of Evita 4 Critical Care Ventilator. Continuous (Respirator) on . The reason FDA records is "audible power failure alarm was not enunciated at the required volume". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2326-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Evita 4 Critical Care Ventilator. Continuous (Respirator)
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
audible power failure alarm was not enunciated at the required volume
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The recalling firm issued an URGENT MEDICAL DEVICE RECALL letter dated 7/2/08 informing the consignees of the problem and the need to have the circuit board replaced.
Quantity in commerce
83 units
Distribution
The products were shipped to medical facilities in AZ, CA, CO, FL, GA, IL, KY, MA, MI, MS, NJ, NY, PA, TN, TX, UT and WA.
Product codes and lots
Catalog number 8411900. Products recalled had the CO2 Carrier Printed Circuit Board (PCB) part number 8306611 replaced between January 2005 and May 2008. The serial number for the affected CO2 Carrier PCBs have a prefix of ARW, ARX, or ARY.