Z-2326-2008
Evita 4 Critical Care Ventilator. Continuous (Respirator)
Draeger, Inc. / initiated 2008-07-02 / Terminated / root cause: manufacturing process
Draeger Medical, Inc. began a recall of Evita 4 Critical Care Ventilator. Continuous (Respirator) on . The reason FDA records is "audible power failure alarm was not enunciated at the required volume". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2326-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger, Inc.
- Product
- Evita 4 Critical Care Ventilator. Continuous (Respirator)
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- audible power failure alarm was not enunciated at the required volume
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The recalling firm issued an URGENT MEDICAL DEVICE RECALL letter dated 7/2/08 informing the consignees of the problem and the need to have the circuit board replaced.
- Quantity in commerce
- 83 units
- Distribution
- The products were shipped to medical facilities in AZ, CA, CO, FL, GA, IL, KY, MA, MI, MS, NJ, NY, PA, TN, TX, UT and WA.
- Product codes and lots
- Catalog number 8411900. Products recalled had the CO2 Carrier Printed Circuit Board (PCB) part number 8306611 replaced between January 2005 and May 2008. The serial number for the affected CO2 Carrier PCBs have a prefix of ARW, ARX, or ARY.