Z-2330-2008

Propper short gas-chex EO sterilization indicators, 250 strips per box, Reorder No. 269001, Propper...

Propper Mfg Co Inc / initiated 2008-02-07 / Terminated / root cause: manufacturing process

Propper Mfg Co Inc began a recall of Propper short gas-chex EO sterilization indicators, 250 strips per box, Reorder No. 269001, Propper Manufacturing Co., Inc., Long Island City, N.Y. 11101 on . The reason FDA records is "The firm discovered a production specification discrepancy that may cause the indicators to show an inaccurate result". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2330-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Propper Mfg Co Inc
Product
Propper short gas-chex EO sterilization indicators, 250 strips per box, Reorder No. 269001, Propper Manufacturing Co., Inc., Long Island City, N.Y. 11101
Product code
JOJ Indicator, Physical/Chemical Sterilization Process
Reason for recall
The firm discovered a production specification discrepancy that may cause the indicators to show an inaccurate result.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
On 2/07/08, Propper Mfg. Co. sent URGENT: DEVICE RECALL letters (dated 2/07/08) to the direct consignees, informing them of the recall of the affected lots of Gas-Chex EO Indicator Strips. The letters also instructed them to: (1) check their inventory to determine if they have the recalled lots in stock; (2) discontinue use of the affected lots; (3) reprocess any unused packs that have been processed with the affected lots of Gas-Chex EO Indicator Strips with non-recalled lots or an alternative product from another supplier; (4) return the affected lots to Propper Manufacturing; (5) complete the enclosed form and return via fax to Propper Mfg. Propper Mfg. will send a postage-paid label and instruction for return of the product, and will ship replacement product as soon as possible. Propper also recommended that consignees review the risk associated with items that have been monitored with the subject chemical indicators although a sterilization cycle failure is unlikely due to equipment controls and other monitoring methods. For questions, please contact Propper at 1-800-832-4300 during regular business hours (9 am to 4:30 pm).
Quantity in commerce
1757 cases (7030 boxes)
Distribution
Nationwide and Puerto Rico
Product codes and lots
Reorder No. 269001; Lots 6081, 6043, 6021, 7042, and 7013