Z-2138-2008
GEM 2753 Microvascular Anastomotic Coupler, 2.5 mm, REF GEM2753/I, P/N 5111-00250-010, Rx only, Irradiation...
SYNOVIS MICRO COMPANIES ALLIANCE, Inc. (a subsidiary of Baxter International Inc.) / initiated 2008-05-14 / Terminated / root cause: design
Synovis Micro Companies Alliance Inc began a recall of GEM 2753 Microvascular Anastomotic Coupler, 2.5 mm, REF GEM2753/I, P/N 5111-00250-010, Rx only, Irradiation Sterilized This is a microsurgical coupler used in... on . The reason FDA records is "Coupler rings may slip out of the delivery toolcaused by jaws larger than specification". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2138-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SYNOVIS MICRO COMPANIES ALLIANCE, Inc. (a subsidiary of Baxter International Inc.)
- Product
- GEM 2753 Microvascular Anastomotic Coupler, 2.5 mm, REF GEM2753/I, P/N 5111-00250-010, Rx only, Irradiation Sterilized This is a microsurgical coupler used in the anastomosis of veins and arteries normally encountered in microsurgical vascular reconstructive procedures.
- Product code
- MVR Device, Anastomotic, Microvascular
- Reason for recall
- Coupler rings may slip out of the delivery toolcaused by jaws larger than specification.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Urgent Recall Notification was sent via Fed Ex on May 15, 2008 to 432 hospitals. The letters were addressed to the attention of the Risk Manager. The recall notification instructed the hospital to immediately discontinue use, examine their inventory and, if found, to remove the product and quarantine it for return. A second notification was sent on July 21, 2008 to non-responding consignees.
- Quantity in commerce
- 3,812 units
- Distribution
- Worldwide Distribution --- including USA and countries of Canada, Japan, UAE, Korea, South Africa, Australia, Taiwan, Norway, Netherlands, Sweden, Greece, Norway, Turkey, Portugal, Taiwan, Denmark, Italy, Germany, Japan, Spain, Austria, and Great Britain.
- Product codes and lots
- Lot numbers: 1808, 1894, 1907, 1923, 1953, 1954, 7139, 7211, 7354, 8998, 8999, 11591, 12364, 15244, 15280, 15281, 18824, 18912, 19294, 19413, 19954, 20263, 20689, 21345, 21515, 21560, 21977, 22556, and 23047