GE Healthcare Innova 3100 IQ Digital Fluoroscopic Imaging System (Cardiovascular imaging system). The Innova...
Datex - Ohmeda, Inc began a recall of GE Healthcare Innova 3100 IQ Digital Fluoroscopic Imaging System (Cardiovascular imaging system). The Innova systems are indicated for use in generating... on . The reason FDA records is "User was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1519-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Innova 3100 IQ Digital Fluoroscopic Imaging System (Cardiovascular imaging system). The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy of cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. These devices are not intended for mammography applications.
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- User was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to incorrect overall dose management, resulting in unnecessary exposure and can be a contributing factor for skin injury. The alarm can be heard in the control room.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- GE Field Engineers are visiting all affected customers to apply updated software. For assistance contact Datax-Ohmeda, Inc., at 1-608-221-1551.
- Quantity in commerce
- 132
- Distribution
- within US: AL, AR, AZ, CA,CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico. OUS to include: Algeria, Australia, Belgium, Brazil, Canada, China, Columbia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Korea, Kuwait, Latvia, Malaysia, Mexico, New Zealand, Norway, Russia, Saudi Arabia, Singapore, Spain, Switzerland, Syria, Taiwan, Turkey, and United Kingdom
- Product codes and lots
- serial numbers: 00000007C20050, 00000061993GE2, 00000464046BU2, 00000464088BU4, 00000464127BU0, 00000477752BU0, 00000477766BU0, 00000477779BU3, 00000477793BU4, 00000477807BU2, 00000477808BU0, 00000477816BU3, 00000477827BU0, 00000477833BU8, 00000477840BU3, 00000479549BU8, 00000479577BU9, 00000485033BU5, 00000489690BU8, 00000489717BU9, 00000490496BU7, 00000491059BU2, 00000492423BU9, 00000493965BU8, 00000497494BU5, 00000498674BU1, 00000498692BU3, 00000498707BU9, 00000498882BU0, 00000498888BU7, 00000499681BU5, 00000499701BU1, 00000499702BU9, 00000499711BU0, 00000499720BU1, 00000499728BU4, 00000499936BU3, 00000505083BU6, 00000505109BU9, 00000509176BU4, 00000509205BU1, 00000509236BU6, 00000509237BU4, 00000509242BU4, 00000509243BU2, 00000509246BU5, 00000509252BU3, 00000509253BU1, 00000510040BU9, 00000511613BU2, 00000511672BU8, 00000511684BU3, 00000511685BU0, 00000511692BU6, 00000512738BU6, 00000512755BU5, 00000512825BU1, 00000512826BU9, 00000512829BU3, 00000512831BU9, 00000512834BU3, 00000512840BU0, 00000512866BU5, 00000512884BU8, 00000512889BU7, 00000512890BU5, 00000512897BU0, 00000512905BU1, 00000513093BU5, 00000513107BU3, 00000513112BU3, 00000513128BU9, 00000513172BU7, 00000513175BU0, 00000513183BU4, 00000513184BU2, 00000513189BU1, 00000513205BU5, 00000513267BU5, 00000513277BU4, 00000513278BU2, 00000513523BU1, 00000513527BU2, 00000513536BU3, 00000513544BU7, 00000513548BU8, 00000513549BU6, 00000513551BU2, 00000513570BU2, 00000513571BU0, 00000513576BU9, 00000513589BU2, 00000515118BU8, 00000518728BU1, 00000519107BU7, 00000519126BU7, 00000519153BU1, 00000520046BU4, 00000521647BU8, 00000526236BU5, 00000526308BU2, 00000526309BU0, 00000526348BU8, 00000526382BU7, 00000526383BU5, 00000526384BU3, 00000526386BU8, 00000526559BU0, 00000528646BU3, 00000529467BU3, 00000529510BU0, 00000529664BU5, 00000529681BU9, 00000530226BU0, 00000534470BU0, 00000534725BU7, 00000538734BU5, 00000538740BU2, 00000543268BU7, 00000545045BU7, 00000545590BU2, 00000546170BU2, 00000546434BU2, 00000546573BU7, 00000547027BU3, 00000547211BU3, 00000549174BU1, 00000772852-02, 00002UA6251HCZ, 00002UA6420FDN. systems id: 2667313, 82416040018, 82416130018.