Fabius Trio Gas machine, anesthesia or analgesia. Catalog number 8606000. The product is used in operating...
Draeger Medical, Inc. began a recall of Fabius Trio Gas machine, anesthesia or analgesia. Catalog number 8606000. The product is used in operating rooms and ER, for patient intubation on . The reason FDA records is "The parent company became aware of instances where the lower side rail of the frame of the machine broke inwards when being moved across a threshold; resulting in the machine being unstable". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2338-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger, Inc.
- Product
- Fabius Trio Gas machine, anesthesia or analgesia. Catalog number 8606000. The product is used in operating rooms and ER, for patient intubation.
- Product code
- BSZ Gas-Machine, Anesthesia
- Reason for recall
- The parent company became aware of instances where the lower side rail of the frame of the machine broke inwards when being moved across a threshold; resulting in the machine being unstable.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm has issued a MEDICAL DEVICE RECALL letter dated 7/7/08 to its customers via certified mail. The letter informs them of the problem and to exercise care when using the device until it has been retrofitted. Contact Drager Medical at 1-800-543-5047, #4 for assistance.
- Quantity in commerce
- 111 units