Z-2344-2008

PLV-100 Portable Lifecare Ventilator Life support ventilation The device is intended to mechanically control...

Respironics, Inc. / initiated 2008-07-08 / Terminated

Respironics, Inc. began a recall of PLV-100 Portable Lifecare Ventilator Life support ventilation The device is intended to mechanically control or assist patient breathing by delivering a... on . The reason FDA records is "Certain PLV-100 ventilators may not trigger a signal to activate third-party remote alarms or nurse call systems upon device failure". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2344-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Respironics, Inc.
Product
PLV-100 Portable Lifecare Ventilator Life support ventilation The device is intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Life support ventilation
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Certain PLV-100 ventilators may not trigger a signal to activate third-party remote alarms or nurse call systems upon device failure.
FDA root cause
Finished device change control
Action
The recalling firm issued an URGENT MEDICAL DEVICE RECALL letter dated 7/8/08. The letter informed the customer of the problem and the need to contact the firm for a replacement. Contact Respironics Customer Service at 1-877-387-331 for assistance.
Quantity in commerce
197
Distribution
Worldwide Distribution ---USA: NC, IL, NJ, MO, TX , OH, IA, and OK and countries of Japan, Denmark, Korea, Australia, Finland, Canada, Argentina, New Zealand, Hong Kong, and India.
Product codes and lots
Model/Catalog number PLV-100. Circuit board part numbers 1034786 version 5 or 1036038 version 3