Z-2347-2008

KyphX Osteo Introducer System with blunt tip introducer stylet (as contained in the KyphoPak Tray, First...

Medtronic, Inc. / initiated 2008-06-23 / Terminated / root cause: design

Kyphon Inc began a recall of KyphX Osteo Introducer System with blunt tip introducer stylet (as contained in the KyphoPak Tray, First Fracture, 15/3 Xpander), catalog number KPT1502... on . The reason FDA records is "Some products may contain the incorrect introducer stylet. If the treating physician is unfamiliar with the use and design of the diamond tip stylet, there is a possibility of serious injuring occurring with the use of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2347-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
KyphX Osteo Introducer System with blunt tip introducer stylet (as contained in the KyphoPak Tray, First Fracture, 15/3 Xpander), catalog number KPT1502, manufactured by Medtronic Spine LLC. Device is used to access vertebral body during balloon kyphoplasty, a procedure designed to reduce and stabilize vertebral body compression fractures (VCFS).
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Some products may contain the incorrect introducer stylet. If the treating physician is unfamiliar with the use and design of the diamond tip stylet, there is a possibility of serious injuring occurring with the use of the diamond tip stylet during balloon kyphoplasty procedure. Other risks include delay to surgery and inconvenience to patient.
Root cause tag
design
FDA root cause
Mixed-up of materials/components
Action
Recall initiated on June 23, 2008. A recall letter and reconciliation form was sent out via e-mail to sales representatives for hand delivery to US hospital consignees. The consignees were asked to review the letter and complete/return the reconciliation form to the firm. If you have any questions contact Kyphon, Inc. at 1-408-548-6740.
Quantity in commerce
310 units
Distribution
Class II Recall - Nationwide Distribution.
Product codes and lots
Lot code J8052810