Zimmer Inc. began a recall of Zimmer Patient Helper Reinforce Overhead Bar for Hill-Rom Versa Care beds; Zimmer Inc., Warsaw, The product is a L shaped bar that is used with hospital beds... on . The reason FDA records is "If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2297-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Patient Helper Reinforce Overhead Bar for Hill-Rom Versa Care beds; Zimmer Inc., Warsaw, The product is a L shaped bar that is used with hospital beds set up for traction. The bar allows mounting of a trapeze to help the patient pull themselves up when in the hospital bed. IN; Prod. Nos. 00-2700-965-02 and 00-0965-001-00.
- Product code
- IKX Aid, Transfer
- Reason for recall
- If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were notified via recall letter dated 7/7/08 to discontinue use of these devices until the firm visits and replaces the devices. Contact Zimmer at 1-800-321-5533 for assistance.
- Quantity in commerce
- 645 of all systems
- Distribution
- Nationwide and Canada.
- Product codes and lots
- All units shipped between 7/11/07 and 5/19/08.