Z-2297-2008

Zimmer Patient Helper Reinforce Overhead Bar for Hill-Rom Versa Care beds; Zimmer Inc., Warsaw, The product...

Zimmer GmbH / initiated 2008-07-07 / Terminated / root cause: design

Zimmer Inc. began a recall of Zimmer Patient Helper Reinforce Overhead Bar for Hill-Rom Versa Care beds; Zimmer Inc., Warsaw, The product is a L shaped bar that is used with hospital beds... on . The reason FDA records is "If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2297-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Patient Helper Reinforce Overhead Bar for Hill-Rom Versa Care beds; Zimmer Inc., Warsaw, The product is a L shaped bar that is used with hospital beds set up for traction. The bar allows mounting of a trapeze to help the patient pull themselves up when in the hospital bed. IN; Prod. Nos. 00-2700-965-02 and 00-0965-001-00.
Product code
IKX Aid, Transfer
Reason for recall
If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were notified via recall letter dated 7/7/08 to discontinue use of these devices until the firm visits and replaces the devices. Contact Zimmer at 1-800-321-5533 for assistance.
Quantity in commerce
645 of all systems
Distribution
Nationwide and Canada.
Product codes and lots
All units shipped between 7/11/07 and 5/19/08.