Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 28 mm diameter, neck length 0 mm, sterile, Zimmer, Warsaw...
Zimmer Inc. began a recall of Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 28 mm diameter, neck length 0 mm, sterile, Zimmer, Warsaw, IN; REF 00-6418-028-02 and 00-6428-028-02. The... on . The reason FDA records is "When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0300-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 28 mm diameter, neck length 0 mm, sterile, Zimmer, Warsaw, IN; REF 00-6418-028-02 and 00-6428-028-02. The Alumina Ceramic Femoral Heads are modular components used in total hip arthroplasty and indicated for patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur; patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis; patients suffering from disability due to previous fusion; patients with previously failed endoprostheses and/or total hip components in the operative extremity; and patients with acute neck fractures.
- Product code
- LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
- Reason for recall
- When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm issued Product Compatibility Update # 13, dated 6/11/08, to consignees informing them which products are not compatible, and thus should not be used in combination.
- Quantity in commerce
- 49,474 of all products
- Distribution
- Worldwide Distribution --- including USA and countries of Austria, Australia, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Dubai, England, Finland, France, Germany, Greece, Hungary, India, Israel, Italy, Korea, Lebanon, Mexico, Netherlands, New Zealand, Poland, Romania, Singapore, Slovakia, Spain, Switzerland and Ukraine.
- Product codes and lots
- All lots.