Z-0302-2009

Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 32 mm diameter, neck length -3.5 mm, sterile, Zimmer...

Zimmer GmbH / initiated 2008-06-11 / Terminated / root cause: design

Zimmer Inc. began a recall of Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 32 mm diameter, neck length -3.5 mm, sterile, Zimmer, Warsaw, IN; REF 00-6418-032-01 and 00-6428-032-01. The... on . The reason FDA records is "When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0302-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 32 mm diameter, neck length -3.5 mm, sterile, Zimmer, Warsaw, IN; REF 00-6418-032-01 and 00-6428-032-01. The Alumina Ceramic Femoral Heads are modular components used in total hip arthroplasty and indicated for patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur; patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis; patients suffering from disability due to previous fusion; patients with previously failed endoprostheses and/or total hip components in the operative extremity; and patients with acute neck fractures.
Product code
LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Reason for recall
When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging.
Root cause tag
design
FDA root cause
Device Design
Action
The firm issued Product Compatibility Update # 13, dated 6/11/08, to consignees informing them which products are not compatible, and thus should not be used in combination.
Quantity in commerce
49,474 of all products
Distribution
Worldwide Distribution --- including USA and countries of Austria, Australia, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Dubai, England, Finland, France, Germany, Greece, Hungary, India, Israel, Italy, Korea, Lebanon, Mexico, Netherlands, New Zealand, Poland, Romania, Singapore, Slovakia, Spain, Switzerland and Ukraine.
Product codes and lots
All lots.