Scorpio Series 7000 Tibial Impactor/Extractor, Catalog number 3770-000 Non Sterile Howmedica Osteonics Corp...
Stryker Howmedica Osteonics Corp. began a recall of Scorpio Series 7000 Tibial Impactor/Extractor, Catalog number 3770-000 Non Sterile Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 The... on . The reason FDA records is "Stryker Orthopaedics became aware that there is a potential inability to release the Scorpio Series 7000 Tibial Impactor/Extractor from a tibial baseplate during use". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0115-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Scorpio Series 7000 Tibial Impactor/Extractor, Catalog number 3770-000 Non Sterile Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 The Impactor/Extractor is used for insertion and removal of a tibial baseplate during total knee arthroplasty.
- Product code
- HWA Impactor
- Reason for recall
- Stryker Orthopaedics became aware that there is a potential inability to release the Scorpio Series 7000 Tibial Impactor/Extractor from a tibial baseplate during use.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- US recall notification (product correction) letters were send out by Federal Express with return receipt on July 2 and 9, 2008. Advisory notices were sent to foreign customers on July 2, 2008. Both letters include a Product Correction Bulletin that instructs all used to lubricate the divide with an approved medical lubricant, especially the extractor shaft and threads. It is recommended that all moving parts be evaluated for functionality and useful service prior to each surgery and instructions on how to perform this process. Contact Stryker Howmedica Osteonics Corp. at 1-201-831-5918 for assistance.
- Quantity in commerce
- 5126
- Distribution
- Worldwide including the US and Asia Pacific, Canada, Europe, Middle East, Africa, Japan and Latin America.
- Product codes and lots
- All Lot Codes