Z-0165-2009
Roche/Hitachi Modular E Module immunoassay analyzer, Roche Diagnostics, Ind., IN; GMMI Nos. 04998642001 and...
Roche Diagnostics GmbH / initiated 2008-06-24 / Terminated / root cause: software
Roche Diagnostics Corp. began a recall of Roche/Hitachi Modular E Module immunoassay analyzer, Roche Diagnostics, Ind., IN; GMMI Nos. 04998642001 and 03617505001 on . The reason FDA records is "A software bug may result in pipetting from an incorrect reagent pack and/or assigning calibration curve parameters incorrectly". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0165-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche/Hitachi Modular E Module immunoassay analyzer, Roche Diagnostics, Ind., IN; GMMI Nos. 04998642001 and 03617505001.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- A software bug may result in pipetting from an incorrect reagent pack and/or assigning calibration curve parameters incorrectly.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- The consignees were notified of the problem via letter dated 6/24/08 and were provided instructions for a workaround.
- Distribution
- Nationwide Distribution --- including states of Florida, New Jersey and North Carolina.
- Product codes and lots
- All units with software version 08-02.