Z-2194-2010
The Philips EasyDiagnost Eleva with DRF. Philips Medical Systems multifunctional x-ray system
Philips North America LLC / initiated 2008-07-09 / Terminated
Philips Medical Systems North America Co. Phillips began a recall of The Philips EasyDiagnost Eleva with DRF. Philips Medical Systems multifunctional x-ray system on . The reason FDA records is "Potential for bright artifacts to appear on the patient image from a previous exposure". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2194-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- The Philips EasyDiagnost Eleva with DRF. Philips Medical Systems multifunctional x-ray system.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- Potential for bright artifacts to appear on the patient image from a previous exposure.
- FDA root cause
- Other
- Action
- On July 9, 2008, Philip Medical System began sending out the URGENT-Device Correction letter to their consignees. This letter describes the recalled product as EasyDiagnost Eleva with DRF system with SW release 1.0. The firm informs the consignees that under special circumstances there is a small possibility that bright artifacts appear on a patient image from a previous exposure. The consignees are advised to take care about the X-ray field limitation. If the detector is not affected by direct radiation, no problem will occur. If direct radiation reaches the detector, the collimation should not be too tight, such that the area of direct radiation is larger than 1.2 inch x 1.2 inch (or 30 mm x 30 mm). The minimum area is needed to detect and prevent memory artifacts. The firm will upgrade any affected products with the Field Change Order (FCO)70600031 and it will be implemented before the end of December 2008. The upgrade software is 1.0.3. The correction would be free of charge and the consignees should contact Philips Call Center at 800-772-9377, #5, #2.
- Quantity in commerce
- 7
- Distribution
- Product was distributed in AZ, CA, FL and WA.
- Product codes and lots
- Serial Numbers 353320, 362600, 362601, 364209, 364210, 375245 and 375850.