Z-2350-2008

Steris Small Renaissance Eagle 3000 -- Remanufactured Gravity Models 3011 and 3021 and Prevacuum Models 3013...

STERIS Corporation / initiated 2008-03-06 / Terminated / root cause: manufacturing process

Steris Corp began a recall of Steris Small Renaissance Eagle 3000 -- Remanufactured Gravity Models 3011 and 3021 and Prevacuum Models 3013 and 3023. Products are designed for efficient... on . The reason FDA records is "The firm discovered that some Eagle 3000 units may have a door post pin missing which could cause incomplete or no engagement of the radial arms into the lock rails of the door. The sterilizer is designed with switches...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2350-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
Steris Small Renaissance Eagle 3000 -- Remanufactured Gravity Models 3011 and 3021 and Prevacuum Models 3013 and 3023. Products are designed for efficient sterilization of non-porous and porous, heat and moisture-stable materials used in healthcare facilities.
Product code
FLE Sterilizer, Steam
Reason for recall
The firm discovered that some Eagle 3000 units may have a door post pin missing which could cause incomplete or no engagement of the radial arms into the lock rails of the door. The sterilizer is designed with switches that prevent cycle initiation if the door is not locked. However, in the event the door switches are not properly adjusted they may falsely indicate that the door is locked even t
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
On March 5, 2008, Steris issued an Urgent Product Correction. Beginning March 6, 2008 customers were notified by telephone. The firm instructed their customers to immediately discontinue using the small Renaissance Eagle 3000 steam sterilizer(s) until inspected and corrected by a service technician. Steris Service technicians were dispatched to affected customers to inspect and correct the sterilizers as appropriate. If you have any questions contact Holly Wright Lee at (404) 392-7019 or 1-800-548-4873 ext. 27019.
Quantity in commerce
148 units
Distribution
Class II Recall - Worldwide Distribution --- USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KY, LA, MD, MI, MN, MO, MT, NC, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WI, and WY, and countries of Canada and Mexico.
Product codes and lots
Gravity Models 3011 & 3021 and Prevaccuum Models 3013 and 3023.