Z-2367-2008

12ml Control Syringe, REF/CAT No.: CCS601 Version B, Sterile R, Merit Medical Systems, Inc., South Jordan...

Merit Medical Systems, Inc. / initiated 2008-07-09 / Terminated / root cause: manufacturing process

Merit Medical Systems, Inc. began a recall of 12ml Control Syringe, REF/CAT No.: CCS601 Version B, Sterile R, Merit Medical Systems, Inc., South Jordan, Utah 84095. The product is used for injections of... on . The reason FDA records is "Control Syringes may be non-sterile due to holes in the packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2367-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
12ml Control Syringe, REF/CAT No.: CCS601 Version B, Sterile R, Merit Medical Systems, Inc., South Jordan, Utah 84095. The product is used for injections of contrast medium during Coronary or Peripheral Angiography and Angioplasty.
Product code
FMF Syringe, Piston
Reason for recall
Control Syringes may be non-sterile due to holes in the packaging.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Urgent Product Recall letters dated 07/09/08, were faxed on 07/09/2008 to consignees/sales reps. The letter instructed consignees/sales reps to contact their customers by phone or visit to provide recall information and instructions. Customers were told to immediately discontinue use and to quarantine all affected products. A Product Retrieval Form was provided that was to be signed by the site representative. Contact Merit Medical at 1-801-208-4408 for assistance.
Quantity in commerce
679 units
Distribution
CA, FL, GA, ID, IL, IN, KS, LA, MO, MT, OH, PA, TN, and WA. Foreign distribution to Japan.
Product codes and lots
Lot Number: F620653