Z-2369-2008
Philips Telemonitoring System Software (M3811B) used with Philips Telemonitoring System (M3810B) Revision...
Philips North America LLC / initiated 2008-07-22 / Terminated / root cause: software
Philips Medical Systems began a recall of Philips Telemonitoring System Software (M3811B) used with Philips Telemonitoring System (M3810B) Revision: B.02.07 Prescriptive medical device used to... on . The reason FDA records is "Software: If Weight Limits edited, values will return to default values that were deleted". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2369-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Telemonitoring System Software (M3811B) used with Philips Telemonitoring System (M3810B) Revision: B.02.07 Prescriptive medical device used to automatically collect and transmit medical information (weight, blood pressure, non-diagnostic ECG) over phone lines between provider and patient.
- Product code
- DQA Oximeter
- Reason for recall
- Software: If Weight Limits edited, values will return to default values that were deleted
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Affected customers will receive the Urgent Medical Device Correction notification on 7/22/08 via certified mail. Delivery receipt will be tracked. The Urgent Medical Device Correction notification instructs customers to verify that daily and weekly weight limits are set or not set as intended while they await the correction, which will come in the form of a new version of software and will simultaneously update all systems. Philips has corrected the software defect and is in the process of validating the correction and releasing the corrected software as part of a scheduled release. If further information or support is needed, customers should contact the Telehealth Customer Service Line at 1-866-246-7316.
- Quantity in commerce
- 49 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Software Revision: B.02.07